Standardization of body composition parameters between GE Lunar iDXA and Hologic Horizon A and their clinical impact
Colin Vendrami,
Guillaume Gatineau,
Elena Gonzalez Rodriguez
et al.
Abstract:Body composition (BC) measured by dual X-ray absorptiometry (DXA) differs between devices. We aimed to compare regional and total BC measurements assessed by the Hologic Horizon A™ and the GE Lunar iDXA™ devices; to determine device-specific calibration equations for each BC parameter; and to assess the impact of this standardization procedure on the assessment of sarcopenia, lipedema, obesity and cardiovascular risk with DXA. A total of 926 postmenopausal women (aged 72.9 ± 6.9 years, height 160.3 ± 6.6 cm, w… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.