Study on calibrating the quadrupole mass spectrometers with gas mixture

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
13
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
5
1

Relationship

0
6

Authors

Journals

citations
Cited by 12 publications
(13 citation statements)
references
References 22 publications
0
13
0
Order By: Relevance
“…In general, for the calibration of pure gases, the most intense signal (base peak) can be selected, but for complex mixtures, the base peaks overlap with the peaks (spectral interferences) of other gases. 11,18,20,22 For the choice of masses, we preferred to refer to the library provided by the MASsoft software (for Hiden analyzers) because the values can differ based on the machines. Karlegärd et al 15 and Le et al 11 report that the more fragments used in the calibration matrix, the more complex the method, the more accurate the analysis, but the longer the measurement time.…”
Section: Determination Of Interferences Using Databases Massoft and Nistmentioning
confidence: 99%
See 3 more Smart Citations
“…In general, for the calibration of pure gases, the most intense signal (base peak) can be selected, but for complex mixtures, the base peaks overlap with the peaks (spectral interferences) of other gases. 11,18,20,22 For the choice of masses, we preferred to refer to the library provided by the MASsoft software (for Hiden analyzers) because the values can differ based on the machines. Karlegärd et al 15 and Le et al 11 report that the more fragments used in the calibration matrix, the more complex the method, the more accurate the analysis, but the longer the measurement time.…”
Section: Determination Of Interferences Using Databases Massoft and Nistmentioning
confidence: 99%
“…24,27 The QMS has low long-term stability; 22,23,28 therefore, its use requires a regular and accurate setup calibration phase. Another difficulty related to the use of the MS is the choice of characteristic peaks (m/z) 11,15,22,27 with as few interferences as possible. For a mixture composed of several gas species, the molecules are ionized and fragmented; these fragments are then separated based on their mass-to-charge ratio (m/z).…”
mentioning
confidence: 99%
See 2 more Smart Citations
“…factors and transmitting e ciencies of a quadrupole mass lter (cf.,Summers, 1969;Lieszkovsky et al, 1990;Jousten, 2008;Malyshev and Middleman, 2008;Yoshida and Arai, 2018; Operating manual of Transpector MPH gas analysis system, PN 074-555-P1C, In con). The relative sensitivities reported for QMS in previous studies indicate more or less variation, but those are within 1 ±~50 % for the above species in most cases (e.g.,Mahaffy et al, 2011;Hirashita et al, 2013;Yoshida et al, 2014;Sun et al, …”
mentioning
confidence: 99%