SU-E-T-137: Dosimetric Validation for Pinnacle, Acuros, AAA, and Brainlab Algorithms with Induced Inhomogenieties

Abstract: Purpose: To compare the dosimetric accuracy of the Eclipse 11.0 Acuros XB and Anisotropic Analytical Algorithm (AAA), Pinnacle‐3 9.2 Collapsed Cone Convolution, and the iPlan 4.1 Monte Carlo (MC) and Pencil Beam (PB) algorithms using measurement as the gold standard. Methods: Ion chamber and diode measurements were taken for 6, 10, and 18 MV beams in a phantom made up of slab densities corresponding to solid water, lung, and bone. The phantom was setup at source‐to‐surface distance of 100 cm, and the field siz… Show more