Abstract:Although integrated EPID images nominally capture the entire dose delivered during an exposure, a small amount of dose is consistently being lost at the end of every exposure. The amount of missing dose is random, depending on the exact beam termination time within a frame. Inclusion of an extra frame at the end of each exposure effectively rectifies the problem, making the EPID more suitable for clinical dosimetry applications. The authors received support from Varian Medical Systems in the form of software a… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.