SU‐E‐T‐171: Missing Dose in Integrated EPID Images

Abstract: Although integrated EPID images nominally capture the entire dose delivered during an exposure, a small amount of dose is consistently being lost at the end of every exposure. The amount of missing dose is random, depending on the exact beam termination time within a frame. Inclusion of an extra frame at the end of each exposure effectively rectifies the problem, making the EPID more suitable for clinical dosimetry applications. The authors received support from Varian Medical Systems in the form of software a… Show more