2012
DOI: 10.1118/1.4735476
|View full text |Cite
|
Sign up to set email alerts
|

SU‐E‐T‐387: Validation of a New System for Patient Specific IMRT QA and Comparison with Other Commerical Systems

Abstract: The Gamma Index values were observed using the MatriXX Multicube, Delta4, and Octavius 4D. The treatment plan included five fields at various gantry angles. Also the gamma index and profiles were calculated for various treatment sites. Delta 4 and the Octavius 4D appears to be quite comparable. Each device has the ability to allow one to verify segmented and composite fields, measure dose profiles and analysis using the Gamma Method. ConclusionsSimilar IMRT QA measurements will be made for more Step-N-Shoot ca… Show more

Help me understand this report

This publication either has no citations yet, or we are still processing them

Set email alert for when this publication receives citations?