2005
DOI: 10.1118/1.1997849
|View full text |Cite
|
Sign up to set email alerts
|

SU‐FF‐T‐177: A Study On Output Factors for Gamma Knife Model B2 Unit with Various Dosimeters

Abstract: Purposes: To verify the output factors (OFs) of Gamma Knife Model B2 by various conventional dosimeters and to determine an effective correction factor for compensating the dosimeter sizes. Method and Materials: The TLDs, radiochromic films, PTW 31002 ion chamber and the diamond detector were used to obtain the OFs, normalized to the 18 mm helmet, for four helmets. The factors were measured in the centre of an 80 mm polystyrene spherical phantom that was positioned at the mechanical center of the machine. The … Show more

Help me understand this report

This publication either has no citations yet, or we are still processing them

Set email alert for when this publication receives citations?

See others like this or search for similar articles