Abstract:Purposes: To verify the output factors (OFs) of Gamma Knife Model B2 by various conventional dosimeters and to determine an effective correction factor for compensating the dosimeter sizes. Method and Materials: The TLDs, radiochromic films, PTW 31002 ion chamber and the diamond detector were used to obtain the OFs, normalized to the 18 mm helmet, for four helmets. The factors were measured in the centre of an 80 mm polystyrene spherical phantom that was positioned at the mechanical center of the machine. The … Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.