Testing Thermostatic Bath End-Scale Stability for Calibration Performance with a Multiple-Sensor Ensemble Using ARIMA, Temporal Stochastics and a Quantum Walker Algorithm

Abstract: Thermostatic bath calibration performance is usually checked for uniformity and stability to serve a wide range of industrial applications. Particularly challenging is the assessment at the limiting specification ends where the sensor system may be less effective in achieving consistency. An ensemble of eight sensors is used to test temperature measurement stability at various topological locations in a thermostatic bath (antifreeze) fluid at −20 °C. Eight streaks of temperature data were collected, and the re… Show more