2008
DOI: 10.1097/01.tp.0000331663.71507.7e
|View full text |Cite
|
Sign up to set email alerts
|

The Emit® 2000 Sirolimus Assay* on the v-Twin®/Viva E® Analyzers

Abstract: Poster Abstracts402 MONDAY Supplement to Transplantation July 27, 2008, Volume 86 Number 2S complexes. The supernatant is then transferred to a photometric cuvette where the enzyme tag is detected using a chromogenic substrate. Results: Analytical and functional sensitivity are 9.5 and 25 ng/mL, respectively. The LoB and LoD are 15.0 and 22.5 ng/mL, respectively, based on a study per the CLSI EP-17A protocol. The method is linear to 500 ng/mL. Repeatability and within-lab reproducibility on patient pools were … Show more