The gel test: sensitivity and specificity for unexpected antibodies to blood group antigens

Abstract: The recently FDA-licensed anti-IgG gel test for pretransfusion antibody detection requires crossover validation before implementation. Six hundred coded samples sent for routine pretransfusion tests were used to compare GEL (ID-MTS, Ortho Diagnostic Systems Inc., Raritan, NJ) with Löw and Messeter's low-ionic-strength saline (LISS). There were 456 GEL-LISS-, 97 GEL+LISS+, 45 GEL-LISS+, and 2 GEL+LISStests. The 144 positive tests involved 157 antibodies; 67 of these (cold auto, anti-M, -Le, etc.) were considere… Show more