Thermal studies on cryocooler thermal masses from 4 K to 300 K for MRI applications
Sushanth R. Shetty,
Ruturaj A. Umaranikar,
S. Kasthurirengan
et al.
Abstract:Gifford–McMahon (GM) cryocoolers are pivotal for maintaining the temperature of MRI magnets, necessitating regular servicing for optimal performance and longevity. This study explores the geometric dimensions and shapes of superconducting magnets utilized in MRI for system cooling. The investigation demonstrates that warm-up time is contingent on heat capacity rather than geometric shapes. As a result, the design of magnets can be customized to specific shapes and sizes in accordance with favorable conditions,… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.