Abstract:UCRL-10030Sturdy silicon transmission counters as thin as 0.0012 in. were produced and used in conjunction with a passive pulse-adder circuit which combined the dE/dx and E pulses so that the nonuniformity and energy straggling in the transmission counter would not decrease the resolution of the energy spectra. Construction and performance of these counters are described.-1-
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.