2022
DOI: 10.1016/j.brachy.2021.08.003
|View full text |Cite
|
Sign up to set email alerts
|

Time-resolved QA and brachytherapy applicator commissioning: Towards the clinical implementation

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

1
4
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
6

Relationship

2
4

Authors

Journals

citations
Cited by 6 publications
(5 citation statements)
references
References 17 publications
1
4
0
Order By: Relevance
“…The median dwell time differences between measured and planned dwell times in the first and last dwell position were 0.1 s and 0.060 s in our study, which coincide with results of Fonseca et al in a recent study [ 34 ]. They described differences less than 0.05 s for all dwell positions except for the first and the last position.…”
Section: Discussionsupporting
confidence: 92%
“…The median dwell time differences between measured and planned dwell times in the first and last dwell position were 0.1 s and 0.060 s in our study, which coincide with results of Fonseca et al in a recent study [ 34 ]. They described differences less than 0.05 s for all dwell positions except for the first and the last position.…”
Section: Discussionsupporting
confidence: 92%
“…For each marker, IP acquisitions were made using the LINAC with the marker in the isocenter, to establish the pixel coordinates corresponding to the center of each marker (Fonseca et al 2021). This calibration procedure is needed, as the IP is also used for applicator QA in our clinic (Fonseca et al 2022). Therefore, the holder has to be removed periodically, and it is not assured that the markers are in the exact same location every time.…”
Section: Source Tracking Measurementmentioning
confidence: 99%
“…The radiotherapy field, the focus of this study, has seen the development of novel and centre-specific technologies as custom 3D printed applicators for brachytherapy, QA tools or patient-specific phantoms for imaging and dosimetry. (Fonseca et al 2022, Membrive Conejo et al 2021, Tino et al 2022.…”
Section: Introductionmentioning
confidence: 99%