ObjectivesObjectives were to 1) compare self-reported function, dexterity, activity performance, quality of life and community integration of the DEKA Arm to conventional prostheses; and 2) examine differences in outcomes by conventional prosthesis type, terminal device type and by DEKA Arm configuration level.MethodsThis was a two-part study; Part A consisted of in-laboratory training. Part B consisted of home use. Study participants were 23 prosthesis users (mean age = 45 ± 16; 87% male) who completed Part A, and 15 (mean age = 45 ± 18; 87% male) who completed Parts A and B. Outcomes including self-report and performance measures, were collected at Baseline using participants’ personal prostheses and at the End of Parts A and B. Scores were compared using paired t-tests. Wilcoxon signed-rank tests were used to compare outcomes for the full sample, and for the sample stratified by device and terminal device type. Analysis of outcomes by configuration level was performed graphically.ResultsAt the End of Part A activity performance using the DEKA Arm and conventional prosthesis was equivalent, but slower with the DEKA Arm. After Part B, performance using the DEKA Arm surpassed conventional prosthesis scores, and speed of activity completion was equivalent. Participants reported using the DEKA Arm to perform more activities, had less perceived disability, and less difficulty in activities at the End of A and B as compared to Baseline. No differences were observed in dexterity, prosthetic skill, spontaneity, pain, community integration or quality of life. Comparisons stratified by device type revealed similar patterns. Graphic comparisons revealed variations by configuration level.ConclusionParticipants using the DEKA Arm had less perceived disability and more engagement of the prosthesis in everyday tasks, although activity performance was slower. After home use experience, activity performance was improved and activity speed equivalent to using conventional prostheses.