For the assessment of the drug azithromycin (AZ) in biological and pharmaceutical samples, three different types of ion selective electrodes were created: the modified carbon paste electrode (MCPE), coated graphite electrode (CGE), and coated wire electrode (CWE). The electrodes followed the Nernst equation in the range (1.0×10 -7 -1.0×10 -1 , 1.0×10 -6 -1.0×10 -1 and 1.0×10 -5 -1.0×10 -1 )M range, with a slope of (58.355, 57.937, and 56,127) mV/decade and a low detection limit of (5.0339×10 -8 , 5.111×10 -7 and 4.8486×10 -6 ) M for (MCPE, CGE and CWE)electrodes respectively. The suggested sensors additionally demonstrated long-term stability and worked following preparation without significant changes in the potential or slope values, the analysis method is very sensitive and selective, and the t-test and F-test at a 95% confidence level were used to perform the statistics precisely and accurately.