Abstract:Pharmacokinetics studies of domperidone generally analyze plasma matrix samples. The present work aimed to develop and validate a rapid and simple reversed phase-HPLC method for quantifying domperidone in plasma matrices. The chromatographic method implemented: 1. Luna Phenomenex® C18 (250 mm × 4.6 mm i.d; 5 µm) column, 2. isocratic mobile phase mixture of phosphate buffer 0.02 M:acetonitrile (70:30, v/v) with a flow rate of 1 mL/min, 3. UV detection at 285 nm. Domperidone and propranolol hydrochloride (as int… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.