2022
DOI: 10.52775/1810-200x-2021-92-4-39-44
|View full text |Cite
|
Sign up to set email alerts
|

Validation of Dosimetric Comissioning Accuracy of IMRT for Versa HD Linear Accelerator with AAPM TG-119 protocol

Abstract: Purpose: Verification of the IMRT prior to the introduction the new Versa HD (Elekta) Linac into operation using test examples given in the AAPM TG-119 report. Materials and methods: The verification of plans created both on the basis of the dose prescriptions given in the AAPMTG-119 report and on the basis of the constraints used in the NMRRC was made. The plans were verified using a matrix of ionization chambers (MatriXX Evolution) placed in a mini Phantom (IBA Dosimetry), using gamma crite… Show more