Where is the common sense in aortic valve replacement? A review of hemodynamics and sizing of stented tissue valves

Abstract: Heated debates revolve around the hemodynamic performance of stented aortic tissue valves. Because the opening area strongly influences the generation of a pressure gradient over the prosthesis, and the outer diameter determines which valve actually fits into the aortic root, it would seem logical that the valve with the greatest opening area in relation to its outer diameter should allow the best hemodynamic performance. Interestingly, neither of these 2 parameters is reflected by the manufacturing companies'… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

1
47
0

Year Published

2014
2014
2024
2024

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 47 publications
(48 citation statements)
references
References 29 publications
1
47
0
Order By: Relevance
“…The cylindrical end determines the appropriate size of the annulus. The corresponding valve replica end is used to visualize supra-annular placement for optimal sizing, thus helping the surgeon to implant the biggest size possible [as is the case with the newer generation tissue valves, such as the Epic Supra, the Perimount Magna Ease (Edwards Lifesciences), the Mosaic Ultra (Medtronic)] [12]. All described features and promising in vitro data were suggestive of eventually reaching a good haemodynamic performance in vivo.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…The cylindrical end determines the appropriate size of the annulus. The corresponding valve replica end is used to visualize supra-annular placement for optimal sizing, thus helping the surgeon to implant the biggest size possible [as is the case with the newer generation tissue valves, such as the Epic Supra, the Perimount Magna Ease (Edwards Lifesciences), the Mosaic Ultra (Medtronic)] [12]. All described features and promising in vitro data were suggestive of eventually reaching a good haemodynamic performance in vivo.…”
Section: Discussionmentioning
confidence: 99%
“…Owing to a lack of uniformity in tissue valve sizes, proper haemodynamic comparison of different tissue valves may be difficult and even misleading. Therefore, non-randomized unmatched comparisons of the haemodynamic performance of different bioprostheses should be interpreted with caution [12]. In an annulus-size-matched ) PPM was present, respectively.…”
Section: Discussionmentioning
confidence: 99%
“…However, fluidodynamic characteristics tend to change when the bioprosthesis is surgically implanted. Indeed, the in vivo scenario is very complex and, in this context, the performance of the prosthesis may be influenced by many factors, not least sizing strategy (2). Consequently, implanting a prosthesis that meets the patient's cardiac output requirement needs experience and knowledge of the true dimensions of the bioprosthesis -its internal and external diameters -and of tissue annulus diameters.…”
Section: Discussionmentioning
confidence: 99%
“…The in vivo hemodynamic performance of a bioprosthesis imBplanted in aortic position is affected by the characteristics of the prosthesis, such as stent design, type and position of the leaflets, sewing-ring size and the sizing strategy adopted (1,2).…”
Section: Accepted Manuscriptmentioning
confidence: 99%
“…14 A minimal prosthesis gradient is achieved through complete resection of the calcified aortic leaflets and supra-annular implantation of a large prosthesis with a maximum valve opening. 16 Of course, the gradient over the implanted valve depends not only on the surgical technique but also on the type of implanted prosthesis. 17 In addition to those described above, the Bernoulli law explains many other phenomena both inside and outside medicine.…”
Section: Surgical Technique Of Valve Implantationmentioning
confidence: 99%