Background:
In nonrandomized, open-label studies, a transcatheter interatrial shunt device (IASD, Corvia Medical) was associated with lower pulmonary capillary wedge pressure (PCWP), fewer symptoms, and greater quality of life and exercise capacity in patients with heart failure (HF) and midrange or preserved ejection fraction (EF ≥40%). We conducted the first randomized sham-controlled trial to evaluate the IASD in HF with EF ≥40%.
Methods:
REDUCE LAP-HF I (Reduce Elevated Left Atrial Pressure in Patients With Heart Failure) was a phase 2, randomized, parallel-group, blinded multicenter trial in patients with New York Heart Association class III or ambulatory class IV HF, EF ≥40%, exercise PCWP ≥25 mm Hg, and PCWP-right atrial pressure gradient ≥5 mm Hg. Participants were randomized (1:1) to the IASD versus a sham procedure (femoral venous access with intracardiac echocardiography but no IASD placement). The participants and investigators assessing the participants during follow-up were blinded to treatment assignment. The primary effectiveness end point was exercise PCWP at 1 month. The primary safety end point was major adverse cardiac, cerebrovascular, and renal events at 1 month. PCWP during exercise was compared between treatment groups using a mixed-effects repeated measures model analysis of covariance that included data from all available stages of exercise.
Results:
A total of 94 patients were enrolled, of whom 44 met inclusion/exclusion criteria and were randomized to the IASD (n=22) and control (n=22) groups. Mean age was 70±9 years, and 50% were female. At 1 month, the IASD resulted in a greater reduction in PCWP compared with sham control (
P
=0.028 accounting for all stages of exercise). Peak PCWP decreased by 3.5±6.4 mm Hg in the treatment group versus 0.5±5.0 mm Hg in the control group (
P
=0.14). There were no peri-procedural or 1-month major adverse cardiac, cerebrovascular, and renal events in the IASD group and 1 event (worsening renal function) in the control group (
P
=1.0).
Conclusions:
In patients with HF and EF ≥40%, IASD treatment reduces PCWP during exercise. Whether this mechanistic effect will translate into sustained improvements in symptoms and outcomes requires further evaluation.
Clinical Trial Registration:
URL:
https://clinicaltrials.gov
. Unique identifier: NCT02600234.
Variables that predict mortality after ablation, include baseline ejection fraction and NYHA functional class, the number of septal arteries injected with ethanol, post-ablation septal thickness, beta-blocker use, and the number of ablation procedures.
The annual rate of appropriate ICD discharges after ASA is low and less than that reported previously for primary prevention of SCD in HCM. This suggests that ASA is not proarrhythmic. Traditional SCD risk factors did not predict ICD shocks in this cohort.
On a per-patient basis, 64-section multi-detector row CT coronary angiography enables the diagnosis of significant (>or=50%) stenosis in CAD with an accuracy of 89%. On a per-segment and per-vessel basis, diagnostic accuracy is still impaired, primarily by limited spatial resolution.
At heart rates of more than 65 beats per minute, use of two-segment reconstruction improves image quality at multidetector CT coronary angiography but does not significantly affect overall diagnostic accuracy compared with single-segment reconstruction.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.