Objective: To compare the feasibility of continuous positive airway pressure (CPAP) support generated by high flow nasal cannula with conventional CPAP for prevention of reintubation among preterm infants with a birth weight of p1250 g.Study Design: Preterm infants were randomized to CPAP generated via high flow cannula or the Infant Flow Nasal CPAP System (VIASYS, Conshohocken, PA, USA) at extubation. Primary outcome was incidence of reintubation within 7 days. Secondary outcomes included change in oxygen use and frequency of apnea and bradycardias postextubation.Results: Forty neonates were randomized. Twelve of 20 infants randomized to high flow cannula CPAP were reintubated compared to three of 20 using Infant Flow (P ¼ 0.003). The high flow cannula group had increased oxygen use and more apneas and bradycardias postextubation.Conclusions: CPAP delivered by high flow nasal cannula failed to maintain extubation status among preterm infants p1250 g as effectively as Infant Flow CPAP.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.