In general, Cerenkov light can cause a problem in detecting a real light signal that is generated in a fiber-optic radiation sensor using an organic scintillator for dose measurements of high-energy electron beams. In this study, the intensity of Cerenkov light is measured and characterized as functions of incident angles, energy, and field size of the electron beam from a LINAC (linear accelerator). To remove or minimize Cerenkov light, a subtraction method using a background optical fiber is investigated.
Purpose
The purpose of this study was to investigate the feasibility of two‐dimensional (2D) dose distribution deconvolution using convolutional neural networks (CNNs) instead of an analytical approach for an in‐house scintillation detector that has a detector‐interface artifact in the penumbra region.
Methods
Datasets of 2D dose distributions were acquired from a medical linear accelerator of Novalis Tx. The datasets comprise two different sizes of square radiation fields and 13 clinical intensity‐modulated radiation treatment (IMRT) plans. These datasets were divided into two datasets (training and test) to train and validate the developed network, called PenumbraNet, which is a shallow linear CNN. The PenumbraNet was trained to transform the measured dose distribution [M(x, y)] to calculated distribution [D(x, y)] by the treatment planning system. After training of the PenumbraNet was completed, the performance was evaluated using test data, which were 10 × 10 cm2 open field and ten clinical IMRT cases. The corrected dose distribution [C(x, y)] was evaluated against D(x, y) with 2%/2 mm and 3%/3 mm criteria of the gamma index for each field. The M(x, y) and deconvolved dose distribution with the analytically obtained kernel using Wiener filtering [A(x, y)] were also evaluated for comparison. In addition, we compared the performance of the shallow depth of linear PenumbraNet with that of nonlinear PenumbraNet and a deep nonlinear PenumbraNet within the same training epoch.
Results
The mean gamma passing rates were 84.77% and 95.81% with 3%/3 mm gamma criteria for A(x, y) and C(x, y) of the PenumbraNet, respectively. The mean gamma pass rates of nonlinear PenumbraNet and the deep depth of nonlinear PenumbraNet were 96.62%, 93.42% with 3%/3 mm gamma criteria, respectively.
Conclusions
We demonstrated the feasibility of the PenumbraNets for 2D dose distribution deconvolution. The nonlinear PenumbraNet which has the best performance improved the gamma passing rate by 11.85% from the M(x, y) at 3%/3 mm gamma criteria.
Megavoltage computed tomography (MVCT) offers an opportunity for adaptive helical tomotherapy. However, high noise and reduced contrast in the MVCT images due to a decrease in the imaging dose to patients limits its usability. Therefore, we propose an algorithm to improve the image quality of MVCT.
Methods:The proposed algorithm generates kilovoltage CT (kVCT)-like images from MVCT images using a cycle-consistency generative adversarial network (cycleGAN)-based image synthesis model. Data augmentation using an affine transformation was applied to the training data to overcome the lack of data diversity in the network training. The mean absolute error (MAE), root-meansquare error (RMSE), peak signal-to-noise ratio (PSNR), and structural similarity index measure (SSIM) were used to quantify the correction accuracy of the images generated by the proposed algorithm. The proposed method was validated by comparing the images generated with those obtained from conventional and deep learning-based image processing method through nonaugmented datasets. Results: The average MAE, RMSE, PSNR, and SSIM values were 18.91 HU, 69.35 HU, 32.73 dB, and 95.48 using the proposed method, respectively, whereas cycleGAN with non-augmented data showed inferior results (19.88 HU, 70.55 HU, 32.62 dB, 95.19, respectively). The voxel values of the image obtained by the proposed method also indicated similar distributions to those of the kVCT image. The dose-volume histogram of the proposed method was also similar to that of electron density corrected MVCT. Conclusions: The proposed algorithm generates synthetic kVCT images from MVCT images using cycleGAN with small patient datasets. The image quality achieved by the proposed method was correspondingly improved to the level of a kVCT image while maintaining the anatomical structure of an MVCT image. The evaluation of dosimetric effectiveness of the proposed method indicates the applicability of accurate treatment planning in adaptive radiation therapy.
In this study, we fabricated a fiber-optic radiation sensor with an organic scintillator to measure the high-energy photon beam from a clinical linear accelerator, and a two-dimensional fiber-optic sensor array to measure high-resolution and real-time dose distributions for small field radiotherapy dosimetry. The scintillating lights generated from each organic sensor probe embedded and arrayed in a water phantom are guided by 10 m plastic optical fibers to the light-measuring device. Two-dimensional photon beam distributions in a water phantom were measured for photon beams with different energies and field sizes. Also, percent depth dose curves for 6 and 15 MV photon beams were obtained.Index Terms-Fiber-optic sensor, organic scintillator, photon beam, plastic optical fiber.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.