Tibial forces were measured in vivo during the first year after total knee arthroplasty in a 66 kg, 80-year-old man. Forces were measured during activities of daily living, rehabilitation, and exercise. Peak tibial forces recorded during walking increased up to 12 months postoperatively (2.8 times body weight). Tibial forces correlated with increasing speed during treadmill walking. Rising from a chair generated peak forces of 2.6 times body weight. Stair descent generated higher peak forces than stair ascent (3.3 versus 2.9 times body weight, respectively). Exercising on a stair-climbing machine generated forces close to two times body weight whereas stationary bicycling generated even lower forces, near one times body weight. In general, the tibial forces recorded during walking and stair climbing were lower than most predicted values. These measurements can be used to validate in vitro and mathematical models of the knee. This should lead to refined surgical techniques and to enhanced prosthetic designs that will improve patient function, patient quality of life, and longevity of total knee arthroplasty implants.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.