Introduction
The degree of interdialytic weight gain and ultrafiltration may influence anemia results in dialysis. The purpose of this study is to evaluate the utility of a blood volume monitor (BVM) in the management of renal anemia and its ability to avoid the variability of hematocrit (Hct) and hemoglobin values (Hb) depending on plasma volume through a simple method of monitoring the total hemoglobin mass (MtHb).
Methods
Predialysis blood samples for measurement were drawn at both the midweek treatment and the beginning‐of‐the‐dialysis‐week treatment in 30 patients. The MtHb was calculated as MtHb = Vb × Hb, where Vb is the absolute blood volume determined by online dialysate dilution using an online hemodiafiltration machine incorporating a relative BVM.
Findings
The MtHb and the total red cell volume (VRBC) as measured with the bolus method at the starting of the treatment were 540 ± 148 grams and 1544 ± 339 mL, respectively. There were significant differences between the Hb levels and between the hematocrit levels according to the time of dialysis. However, the MtHb remained constant. There was also an excellent correlation between the Hb measurements by the BVM and the blood sampling method (R = 0.89, P value <0.001).
Conclusion
Our study suggests that BVM could be very useful in the management of anemia in dialysis by computing the total Hb mass in clinical practice and may support better and more appropriate assessments of the factors influencing circulating Hb.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.