Background
Patients in intensive care units (ICUs) frequently experience adverse drug events involving intravenous medications (IV-ADEs), which are often preventable.
Objectives
To determine how frequently preventable IV-ADEs in ICUs match the safety features of a programmable infusion pump with safety software (“smart pump”) and to suggest potential improvements in smart-pump design.
Design
Using retrospective medical-record review, we examined preventable IV-ADEs in ICUs before and after 2 hospitals replaced conventional pumps with smart pumps. The smart pumps alerted users when programmed to deliver duplicate infusions or continuous-infusion doses outside hospital-defined ranges.
Participants
4,604 critically ill adults at 1 academic and 1 nonacademic hospital.
Measurements
Preventable IV-ADEs matching smart-pump features and errors involved in preventable IV-ADEs.
Results
Of 100 preventable IV-ADEs identified, 4 involved errors matching smart-pump features. Two occurred before and 2 after smart-pump implementation. Overall, 29% of preventable IV-ADEs involved overdoses; 37%, failures to monitor for potential problems; and 45%, failures to intervene when problems appeared. Error descriptions suggested that expanding smart pumps’ capabilities might enable them to prevent more IV-ADEs.
Conclusion
The smart pumps we evaluated are unlikely to reduce preventable IV-ADEs in ICUs because they address only 4% of them. Expanding smart-pump capabilities might prevent more IV-ADEs.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.