Aim: A reliable, sensitive, HPLC method was developed and validated to simultaneously quantify raloxifene (RLX) and cladrin (CLD). Method: The C18 column was used to analyze RLX and CLD at λmax 285 and 249 nm. The mobile phase was composed of acetonitrile and 35:65% v/v aqueous solution of 0.1% formic acid. Results: The method was linear over the linearity range of 0.078–20 μg/ml, and the limit of detection and limit of quantification for RLX and CLD were 0.191 and 0.228 and 0.581 and 0.69 μg/ml, respectively. Conclusion: In accordance with the International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use guidelines, the developed method is precise and accurate for simultaneous estimation of RLX and CLD with applications in in vitro liver microsomal stability in mice, rabbits, dogs, monkeys and humans.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.