For the detection of pulmonary nodules, the performance of chest tomosynthesis is better, with increased sensitivity especially for nodules smaller than 9 mm, than that of chest radiography.
Observer performance studies are time-consuming tasks, both for the participating observers and for the scientists collecting and analysing the data. A possible way to optimise such studies is to perform them in a completely digital environment. A software tool-ViewDEX (Viewer for Digital Evaluation of X-ray images)-has been developed in Java, enabling it to function on almost any computer. ViewDEX is designed to handle several types of studies, such as visual grading analysis (VGA), image criteria scoring (ICS) and receiver operating characteristics (ROC). The results from each observer are saved in a log file, which can be exported for further analysis in, for example, a special software for analysing ROC results. By using ViewDEX for an ROC experiment, an evaluation rate of approximately 200 images per hour can be achieved, compared to approximately 25 images per hour using hard copy evaluation. The results are obtained within minutes of completion of the viewing. The risk of human errors in the process of data collection and analysis is also minimised. The viewer has been used in a major trial containing approximately 2700 images.
In diagnostic radiology, medical-grade monochrome displays are usually recommended because of their higher luminance. Standard color displays can be used as a less expensive alternative, but have a lower luminance. The aim of the present study was to compare image quality for these two types of displays. Images of a CDRAD contrast-detail phantom were read by four radiologists using a 2-megapixel (MP) color display (143 cd/m 2 maximum luminance) as well as 2-MP (295 cd/m 2 ) and 3-MP monochrome displays. Thirty lumbar spine radiographs were also read by four radiologists using the color and the 2-MP monochrome display in a visual grading analysis (VGA). Very small differences were found between the displays when reading the CDRAD images. The VGA scores were j0.28 for the color and j0.25 for the monochrome display (p=0.24; NS). It thus seems possible to use color displays in diagnostic radiology provided that grayscale adjustment is used.
The study indicates that experienced thoracic radiologists already within the first months of clinical use of chest tomosynthesis are able to take advantage of the new technique in the task of detecting pulmonary nodules.
The results indicate that inexperienced observers can reach a high level of performance regarding nodule detection in tomosynthesis after learning with feedback and that the main problem with chest tomosynthesis is related to the limited depth resolution.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.