Overnight 12 hour tape recordings were made of arterial oxygen saturation (SaO2, pulse oximeter in the beat to beat mode) and abdominal wall breathing movement on 67 healthy, full term infants between the ages of 29 and 54 (median 39) days. The median baseline SaO2 during regular breathing was 99-8% (range 97-0-100%). Fifty four infants (81%) had shortlived episodes during which SaO2 fell to 80% or less (desaturation); the median rate was 0-9 desaturations/hour, and the median duration of each desaturation was 1-2 seconds. The 97th centile value for the duration of all episodes in which SaO2 fell to 680% was 4-0 seconds. The frequency of desaturations was significantly higher, and their duration significantly longer, when the breathing pattern was non-regular rather than regular. The percentage of apnoeic pauses (¢4 seconds in duration) followed by a desaturation was higher during non-regular than regular breathing; it was particularly high during periodic breathing.A knowledge of normal variability of baseline measurements of oxygenation and of the relationship between oxygenation and breathing patterns in infants is essential to the use of pulse oximetry in clinical practice.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.