As an alternative treatment for chronic back pain due to disc degeneration motion preserving techniques such as posterior dynamic stabilization (PDS) has been clinically introduced, with the intention to alter the load transfer and the kinematics at the affected level to delay degeneration. However, up to the present, it remains unclear when a PDS is clinically indicated and how the ideal PDS mechanism should be designed to achieve this goal. Therefore, the objective of this study was to compare different PDS devices against rigid fixation to investigate the biomechanical impact of PDS design on stabilization and load transfer in the treated and adjacent cranial segment. Six human lumbar spine specimens (L3-L5) were tested in a spine loading apparatus. In vitro flexibility testing was performed by applying pure bending moments of 7.5 Nm without and with additional preload of 400 N in the three principal motion planes. Four PDS devices,
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.