“…The linearity and limits of detection (LOD) and quantification (LOQ) were assessed. Solutions from 10 to 2000 ng/mL of 10-HDA, 10-hydroxydecanoic acid, 3-hydroxydecanoic acid, decanedioic acid and 2-dodecenedioic acid [3 replicates; 16 levels (10,30,50,80,100,200, 300, 400, 500, 600, 700, 800, 1000, 1300, 1500, 2000 ng/mL); n = 3 × 16] and solutions from 5-500 ng/mL of decanoic acid and dodecanoic acid were used for the construction of the linear curves [3 replicates; 8 levels (10, 30, 50, 100, 200, 300, 400, 500 ng/mL); n = 3 × 8]. The instrumental detection limits were estimated (LOD = 3.3 * Sa/b, where Sa is the standard deviation of the response of six replicates of the lowest standard and b is the slope of the calibration curve).…”