3D VMAT Verification Based on Monte Carlo Log File Simulation with Experimental Feedback from Film Dosimetry

Abstract: A model based on a specific phantom, called QuAArC, has been designed for the evaluation of planning and verification systems of complex radiotherapy treatments, such as volumetric modulated arc therapy (VMAT). This model uses the high accuracy provided by the Monte Carlo (MC) simulation of log files and allows the experimental feedback from the high spatial resolution of films hosted in QuAArC. This cylindrical phantom was specifically designed to host films rolled at different radial distances able to take i… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
12
0

Year Published

2017
2017
2021
2021

Publication Types

Select...
8

Relationship

0
8

Authors

Journals

citations
Cited by 12 publications
(12 citation statements)
references
References 29 publications
0
12
0
Order By: Relevance
“…Similarly, Barbeiro et al. found log files reconstructed with ~3 times as many control points as the TPS agreed better with film than those reconstructed with the TPS’ discretization …”
Section: Resultsmentioning
confidence: 87%
See 1 more Smart Citation
“…Similarly, Barbeiro et al. found log files reconstructed with ~3 times as many control points as the TPS agreed better with film than those reconstructed with the TPS’ discretization …”
Section: Resultsmentioning
confidence: 87%
“…times as many control points as the TPS agreed better with film than those reconstructed with the TPS' discretization. 26 Using 2-mm dose grid resolution and 1°control point spacing, computational cost for CS was 12 s per control point. This corresponds to 18 min for a 358°arc planned using 4 degree control point spacing or 72 min for a log file plan reconstructed using 1°control point spacing.…”
Section: A Preliminary Validationsmentioning
confidence: 99%
“…Indeed, Monaco TPS provides the MC and Collapsed Cone algorithms coupling. 42 Monaco addressed MC in Kernel evaluation; however, the Therma method of calculation remained. Therefore, Monaco may not be considered as a full implementation of MC for clinical applications.…”
Section: Discussionmentioning
confidence: 99%
“…These attributes have proven useful for a variety of clinical tasks such as dosimetry for total skin electron therapy [9,10] or total body irradiation [11]. RCF is also used in patient specific quality assurance (QA) measurements for delivery techniques such as volumetric modulated arc therapy (VMAT) [12] or stereotactic radiosurgery (SRS) [13,14] where high spatial resolution is necessary and has proven useful as a research tool in phantom and radiobiological studies to verify dosimetry in unconventional geometries that do not easily accommodate conventional dosimeters [15][16][17][18][19][20][21].…”
Section: Introductionmentioning