2021
DOI: 10.1002/mp.14924
|View full text |Cite
|
Sign up to set email alerts
|

8+ Year Performance of the Gamma Knife Perfexion/Icon Patient Positioning System and Possibilities for Preemptive Fault Detection Using Statistical Process Control

Abstract: Background: The large fractional doses, steep dose gradients, and small targets found in intracranial radiosurgery require extremely low beam delivery uncertainty. In the case of Gamma Knife radiosurgery (GKRS), this includes minimizing patient positioning system (PPS) positioning uncertainty. Existing QA techniques are recipe based, and feature point in time pass/fail tolerances. However, modern treatment machines, including the Gamma Knife Perfexion/Icon systems, record extensive internal data in treatment l… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

1
3
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(4 citation statements)
references
References 40 publications
1
3
0
Order By: Relevance
“…These films were then scanned into FilmQAPro 2016 by an Epson 10000XL transmission scanner and the profiles (radiological center) were analyzed in comparison to the location of the pinprick (mechanical center). The Elekta‐provided and calibrated QA diode tool was used prior to treatments for the daily “focus precision test” for over a year to measure the position of the radiological focus compared to the position calibrated into the patient positioning system. These results were recorded and trended over that time, giving an indication of the stability of this agreement consistent with a recommended methodology involving statistical process control 32 …”
Section: Methodssupporting
confidence: 81%
See 2 more Smart Citations
“…These films were then scanned into FilmQAPro 2016 by an Epson 10000XL transmission scanner and the profiles (radiological center) were analyzed in comparison to the location of the pinprick (mechanical center). The Elekta‐provided and calibrated QA diode tool was used prior to treatments for the daily “focus precision test” for over a year to measure the position of the radiological focus compared to the position calibrated into the patient positioning system. These results were recorded and trended over that time, giving an indication of the stability of this agreement consistent with a recommended methodology involving statistical process control 32 …”
Section: Methodssupporting
confidence: 81%
“…This HDMM marker data were retrospectively collected and analyzed. The patient‐specific random uncertainty related to the HDMM data was analyzed versus various patient‐related factors with the statistical analysis to determine potential predictive variables 32 …”
Section: Methodsmentioning
confidence: 99%
See 1 more Smart Citation
“…The patient is connected up to the system by means of a stereotactic helmet that provides treatment coordinates and prevents intra-fraction movements. Accuracy is sub-millimeter, which is ideal for intracranial targets [40].…”
Section: • Tumor Position Reconstruction By Means Of Anmentioning