2023
DOI: 10.1097/phm.0000000000002285
|View full text |Cite
|
Sign up to set email alerts
|

A case study of intrathecal baclofen pump motor shutdown secondary to the effect of the magnetic field created by a personal digital tablet and magnetic cover

Abstract: We present the case of a posttraumatic C6 AIS A tetraplegic patient with spasticity treated with an intrathecal baclofen pump, who noticed a transient increase in his spasticity each time he used a digital tablet (Ipad) protected by a magnetic shell placed on his abdomen. Telemetry confirmed transient motor shutdown responsible for withdrawal symptoms each time the tablet was used. Symptoms resolved after the removal of the protective shell. Effects of magnetic fields like magnetic resonance imaging are known … Show more