A Catastrophic Implant Failure After Total Knee Arthroplasty

Abstract: Wear of metallic surfaces after total knee arthroplasty is uncommon. In this case, we report a catastrophic failure of the knee prosthesis in a 60-year-old male patient who had undergone total knee prosthesis 7 years prior to his admission to our clinic. Genu recurvatum was observed in the affected knee. Surgical exploration of the knee showed wear of the polyethylene insert accompanied by metallic tibial component abrasion of the posterior-cruciate-retaining total knee prosthesis (Maxim, Biomet, Warsaw, IN, U… Show more