A Comparative Effectiveness Analysis of Three Continuous Glucose Monitors

Abstract: The effectiveness and safety of continuous glucose monitors (CGMs) is dependent on their accuracy and reliability. The objective of this study was to compare 3 CGMs in adult and pediatric subjects with type 1 diabetes under closed-loop blood-glucose (BG) control. Twenty-four subjects (12 adults) with type 1 diabetes each participated in one 48-hour closed-loop BG control experiment. Venous plasma glucose (PG) measurements obtained every 15 minutes (4657 values) were paired in time with corresponding CGM glucos… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

8
157
0
4

Year Published

2015
2015
2021
2021

Publication Types

Select...
7
1

Relationship

0
8

Authors

Journals

citations
Cited by 157 publications
(169 citation statements)
references
References 6 publications
8
157
0
4
Order By: Relevance
“…The MARD and PARD results obtained from complete experiments are in agreement with previous reports about the three CGM systems, [2][3][4][5][6][7][8][9][10]12,[16][17][18][19] with the Navigator achieving better performance than the other two systems. Results obtained from core phase evaluation, i.e., during days 2 to 5 when all sensors were used in parallel minus ~24 h runin time, differ only slightly from results for complete experiments, thus the influence from run-in time and different duration of sensor usage was, in this study, less important than expected.…”
Section: Discussionsupporting
confidence: 81%
See 2 more Smart Citations
“…The MARD and PARD results obtained from complete experiments are in agreement with previous reports about the three CGM systems, [2][3][4][5][6][7][8][9][10]12,[16][17][18][19] with the Navigator achieving better performance than the other two systems. Results obtained from core phase evaluation, i.e., during days 2 to 5 when all sensors were used in parallel minus ~24 h runin time, differ only slightly from results for complete experiments, thus the influence from run-in time and different duration of sensor usage was, in this study, less important than expected.…”
Section: Discussionsupporting
confidence: 81%
“…This result is in agreement with other studies incorporating EGA analysis. 3,8,9,12,13,[16][17][18][19]21,22 Only 70% to 80% of paired readings were found to be within the system accuracy limits of ISO 15197:2003. 20 According to ISO 15197, 95% of device readings must be within ±15 mg/dl or ±20% of the paired reference readings below or above 75 mg/dl.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…While implantable sensors have been developed for continuous glucose monitoring, they suffer from limitations in precision and accuracy. 87 Improved sensing capabilities are on the horizon, with influences from lab-on-chip technology, potentially building on miniaturized enzymatic amperometric sensors as described by Senapati et al, where the charge of the captured target molecules is used to block ionic current through an ion-selective membrane 88 . Further advances in decreasing the required sample analysis volume are needed and reducing the size and cost of the sensors.…”
Section: Emergence Of Smart Biofunctional Microneedles With Combined mentioning
confidence: 99%
“…However, it should also be made clear that, while CGM offers many clinical and research advantages, it is an imperfect surrogate for venous blood glucose. Signal stability, physiological differences in interstitial glucose homeostasis, physiological/device lag-times and individual differences are all factors that require consideration when interpreting CGM data [20,21]. Further, since CGM only measures glucose levels, it does not provide any estimation of insulin sensitivity or beta cell function--key information to fully understanding glycaemic responses.…”