1986
DOI: 10.1051/ject/1986182055
|View full text |Cite
|
Sign up to set email alerts
|

A Comparison of a Heparin Coated and a Non-Coated ECC Arterial Line Blood Filter

G.A. Justison,
H.D. Lewis,
D.G. Hamilton
et al.

Abstract: Two groups of five heparin coated (American Bentley AF 1040C) and non-coated (Gish Biomedical, Inc. AFS 40) ECC arterial line filters were compared In Vitro for pressure drop, debubbling time, air separating ability and clinically for blood handling ability. After CO2 flush and following Directions for Use, the debubbling time for the AF 1040C was 178 ± 11 sec. (1 S.D.) and 183 ± 29 sec. for the AFS 40. The preparation times were not significantly different and were confirmed by doppler. The AFS 40 exhibited a… Show more