Abstract:In this paper, we demonstrated the comparison of digital-based CNT and analogue-based filament X-ray sources by taking the X-ray images of ACR mammo phantom for developing the intraoperative specimen X-ray system. The X-ray image of ACR mammo phantom taken at 25 kV 1 mA by filament shows overall the better image quality. However, when X-ray images of phantom were compared at 40 kV 1 mA by both sources, it showed that CNT X-ray source showed better image quality because of Aluminum window.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.