Objective::
This study aimed to establish a simple, reliable, and sensitive method for detecting
ketamine, fluoroketamine, and their metabolites in urine using UPLC-MS/MS.
Methods::
The chromatographic separation was performed on UPLC BEH C18 (50 mm × 2.1 mm,
1.7 μm) at a column temperature of 40°C. The mobile phase consisted of 0.1% formic acid aqueous
solution and acetonitrile, with a flow rate set at 0.4 mL/min, following a specific elution procedure.
A urine sample was treated with acetonitrile, and midazolam was used as an internal standard.
Multiple reaction monitoring was used for quantitative analysis.
Results::
Ketamine, fluoroketamine, norketamine, and 2-norfluoro-ketamine exhibited linearity in
urine (r>0.99) within the concentration range of 5–2000 ng/mL. Intra-day and inter-day precisions
were 9% or less and 12% or less, respectively. The accuracy ranged from 92 to 107%. Mean recoveries
were above 76%. The measured matrix effect was between 85 and 104%.
Conclusion::
This simple, reliable, and sensitive PLC-MS/MS method was successfully developed
to determine ketamine, fluoroketamine, and their metabolite in rat urine.