2016
DOI: 10.1002/acm2.12000
|View full text |Cite
|
Sign up to set email alerts
|

A method for confirming a third‐party assay of I‐125 seeds used for prostate implants

Abstract: The purpose of this work is to describe a method and apparatus that can be used to confirm the source strength of a large number of I‐125 seeds while maintaining sterility, accuracy, reproducibility, and time efficiency. Source strengths ranging from 0.395 to 0.504 U/seed were available for this study. Three different seed configurations were measured: loose, linked, and loaded needles. A third‐party 10% assay (NIST traceable) was provided. A custom stand was built out of aluminum to hold an exposure meter [In… Show more