A reliable HPLC method for the simultaneous estimation of Silodosin and Solifenacin was developed using the quality by design approach. In this study, the main goal was to find a dependable chromatographic system for acceptable component separation. The peaks were sharp within a reasonable run time. For the risk measurement related to the study, critical quality attributes were evaluated.To create an analytical design and a control strategy, the quantitative relationship between critical materials attributes and critical quality attributes was built utilizing the Box-Behnken design. The optimum and ideal mobile phase ratio selected as per the quality by design approach was acetonitrile:methanol:buffer (50:20:30, v/v/v) which has a pH 4.5 flow rate of 1 ml/min. The buffer system used in this mobile phase was potassium dihydrogen orthophosphate. The retention time for Silodosin and Solifenacin determined was 3.626 and 5.754 min. respectively at λmax = 210 nm. The R 2 value calculated for Silodosin and Solifenacin was 0.9994 and 0.9982, respectively. The related standard deviation was less than 2% for the developed method showing good reproducibility and recovery. A simple, precise, and repeatable high-performance liquid chromatography method for regular quality control testing was developed simultaneously by applying a quality by design approach.