Background
Hemostatic devices are now frequently used in femoral artery punctures, and the Angio-Seal (Terumo, Tokyo, Japan) is one of the most commonly used devices for closure of the femoral artery because it provides rapid hemostasis. Although device failure rarely occurs, if the collagen falls into the femoral artery, it may lead to severe limb ischemia. Herein, we describe a case of a novel endovascular technique for the treatment of Angio-Seal arterial closure device failure.
Case presentation
The patient in Case 1 was a 75-year-old man with severe left limb claudication. We used a contralateral antegrade approach and used the Angio-Seal for hemostasis. However, the Angio-Seal collagen and footplate dropped and stopped at the bifurcation of the superficial femoral artery and deep femoral artery. The collagen with the footplate was caught with myocardial biotome forceps (MBF) and pulled into the external iliac artery (EIA). The distal common femoral artery (CFA) was punctured, and we delivered a 10.0- × 80-mm stent (SMART®; Cordis, USA) to the EIA from the ipsilateral sheath. The stent was deployed at the EIA and crushed the collagen. The patient in Case 2 was an 88-year-old man with rest pain in the right limb. The right CFA was punctured using an ipsilateral approach and the Angio-Seal was used for hemostasis. The Angio-Seal collagen with the footplate dropped into the bifurcation of the deep femoral artery. The collagen and footplate were caught with MBF and pulled up to the EIA. The right CFA was punctured and a 10.0- × 60-mm stent (SMART®; Cordis) was delivered from the ipsilateral sheath. The stent was deployed at the EIA and crushed the collagen with the footplate.
Conclusions
MBF were used to grasp the dislodged collagen with the anchor and cover it with a stent at the iliac artery. This may be a useful bailout technique for Angio-Seal dislodgement.