Introduction/Objective:
The current study examines the methodical Quality by Design
(QbD) that facilitated the creation of an easy-to-use, quick, affordable, and stability-indicating reversed-
phase high-performance liquid chromatography (RP-HPLC) technique for the efficient analysis
of aloe-emodin.
Methods:
The chromatographic conditions were optimized with the Design Expert software 11.0
version, i.e., flow rate, buffer concentration, and column temperature.
Results:
The results of the linearity graph show R2 = 0.9988. The LOQ was 0.07949 μg/mL and
the LOD was 0.02623 μg/mL. According to ICH rules, the technique validation parameters were
within the allowed range. Utilizing the Design Expert 11.0 version, the Box–Behnken design experimental
design explains the relationships between flow rate, buffer concentration, and column
temperature at three distinct levels. The responses were monitored: the retention time (Rt), tailing
factor (Tf), and number of theoretical plates (NTPs).
Conclusion:
The suggested approach was appropriate for quantitative determination and may be
used in clinical pharmacokinetic investigations, biopharmaceutics, accredited testing laboratories,
and quality control departments in enterprises.