1982
DOI: 10.1118/1.595124
|View full text |Cite
|
Sign up to set email alerts
|

A practical method to measure the MTF of CT scanners

Abstract: A method has been devised which enables the modulation transfer function (MTF) of computerized tomographic (CT) scanners to be simply and rapidly determined. The method relies upon the measurement of the standard deviation of the pixel values within the image of cyclic bar patterns. The method is shown to produce results which are indistinguishable from those derived by less practical conventional methods.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
129
0

Year Published

2015
2015
2021
2021

Publication Types

Select...
7
2

Relationship

0
9

Authors

Journals

citations
Cited by 174 publications
(129 citation statements)
references
References 0 publications
0
129
0
Order By: Relevance
“…In imaging systems, the mathematical properties of Eq. (1) have been used to characterize the relationship between the contrast transfer function of square wave pattern compared to the sinusoidal varying patterns of the MTF (12) …”
Section: Results and Analysismentioning
confidence: 99%
“…In imaging systems, the mathematical properties of Eq. (1) have been used to characterize the relationship between the contrast transfer function of square wave pattern compared to the sinusoidal varying patterns of the MTF (12) …”
Section: Results and Analysismentioning
confidence: 99%
“…High‐contrast resolution was calculated using the method reported by Droege et al13 In that method, the practical modulation transfer function (pMTF) curve is calculated by measuring the standard deviation of the pixel values in each individual pattern in the cyclic bar pattern image. To assess the spatial resolution quantitatively, 50% and 10% values were calculated from the pMTF curve data.…”
Section: Methodsmentioning
confidence: 99%
“…MTF was calculated by applying the Droege‐Morin (D‐M) method (12) to the line‐pair bar regions in the module. The high resolution gauges in the phantom is made of 2 mm thick aluminum sheet with different width casted into epoxy.…”
Section: Methodsmentioning