Background:
The objective of this study was to assess the safety and effectiveness of the first commercial configuration of a tubeless automated insulin delivery system, Omnipod
®
5, in children (6–13.9 years) and adults (14–70 years) with type 1 diabetes (T1D) in an outpatient setting.
Materials and Methods:
This was a single-arm, multicenter, prospective clinical study. Data were collected over a 14-day standard therapy (ST) phase followed by a 14-day hybrid closed-loop (HCL) phase, where participants (
n
= 36) spent 72 h at each of three prespecified glucose targets (130, 140, and 150 mg/dL, 9 days total) then 5 days with free choice of glucose targets (110–150 mg/dL) using the Omnipod 5. Remote safety monitoring alerts were enabled during the HCL phase. Primary endpoints were difference in time in range (TIR) (70–180 mg/dL) between ST and HCL phases and proportion of participants reporting serious device-related adverse events.
Results:
Mean TIR was significantly higher among children in the free-choice period overall (64.9% ± 12.2%,
P
< 0.01) and when using a 110 mg/dL target (71.2% ± 10.2%,
P
< 0.01), a 130 mg/dL target (61.5% ± 7.7%,
P
< 0.01), and a 140 mg/dL target (64.8% ± 11.6%,
P
< 0.01), and among adults using a 130 mg/dL target (75.1% ± 11.6%,
P
< 0.05), compared to the ST phase (children: 51.0% ± 13.3% and adults: 65.6% ± 15.7%). There were no serious device-related adverse events reported during the HCL phase, nor were there episodes of severe hypoglycemia or diabetic ketoacidosis.
Conclusion:
The Omnipod 5 System was safe and effective when used at glucose targets from 110 to 150 mg/dL for 14 days at home in children and adults with T1D.