Absence of shock therapy related to improper sensing of noise on the defibrillation test during subcutaneous implantable cardioverter–defibrillator implantation: a case report
Abstract:Background
Subcutaneous implantable cardioverter–defibrillator (S-ICD) represents an efficient alternative to transvenous ICD in patients who do not require pacing. The intraoperative defibrillation test (DFT) is recommended during S-ICD implantation to confirm appropriate sensing and successful 65-J termination of induced ventricular fibrillation (VF). However, few cases of oversensing of noise inhibiting therapies have been reported.
Case s… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.