In the present study, we show that CZT detectors can be used in a wide X-ray energy range including low-energy X-rays down to 2.1 keV at which may be extended down to 1.5 keV by additional cooling to . CZT detector was operated in the low energy X-ray band keV to detect the K fluorescence X-rays emitted by low-Z elements, such as Ca, K, Cl, S, and eventually even further P, Si, Al. Additionally, it was demonstrated the fine spectroscopic features of the CZT detector cooled to , which make it suitable for the detection of L fluorescence X-rays emitted by high-Z elements in the keV range, such as Th, Pa, U, Np and Pu, resulting from the decay of actinide radionuclides , , and , respectively. With this cooled CZT detector it was possible to resolve the main L X-ray lines as well as to measure its intensity. The relative intensity obtained for the Pu L X-ray lines , , , , , and was, respectively, , , , , and . The Np L X-ray most intense lines , , , and yielded a relative intensity of , , and , respectively. The overall results are in good agreement with the values found in literature. The principal performance features of the CZT detector were, also studied experimentally or estimated from data tabulated in the literature. Notably, we found a -fold diminution of the electronic noise FWHM temperature sensitivity (from to eV ) and thus of the equivalent noise charge temperature sensitivity (from to ) when the temperature decreases from to . The influence of several properties, such as the charge mobility and lifetime, band gap energy, detector's leakage current and capacitance, on CZT detector performance features was also addressed.