Accuracy of GFR estimating equations combining standardized cystatin C and creatinine assays: a cross-sectional study in Sweden

Abstract: Combining cystatin C and creatinine assays improves GFR estimations with P30 ≥90% in adults. Reporting estimates of both single and combined marker equations in clinical settings makes it possible to assess the validity of the combined equation based on the agreement between the single marker equations.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

6
59
1
1

Year Published

2015
2015
2023
2023

Publication Types

Select...
6

Relationship

1
5

Authors

Journals

citations
Cited by 83 publications
(67 citation statements)
references
References 16 publications
6
59
1
1
Order By: Relevance
“…Mean age was slightly higher in the BIS population at baseline with 80.4 versus 78.5 years in the iohexol population. There were 53% females, 79% on antihypertensive medication, 26% were diabetics, and the mean BSA was 1.81 m 2 in the total BIS population; characteristics of the iohexol population were 43% females, 77% individuals with hypertension, 24% diabetics, and a mean BSA of 1.85 m 2 In Table 2 an overview of the three cysC assays and the two BIS study populations is displayed. …”
Section: Resultsmentioning
confidence: 99%
See 4 more Smart Citations
“…Mean age was slightly higher in the BIS population at baseline with 80.4 versus 78.5 years in the iohexol population. There were 53% females, 79% on antihypertensive medication, 26% were diabetics, and the mean BSA was 1.81 m 2 in the total BIS population; characteristics of the iohexol population were 43% females, 77% individuals with hypertension, 24% diabetics, and a mean BSA of 1.85 m 2 In Table 2 an overview of the three cysC assays and the two BIS study populations is displayed. …”
Section: Resultsmentioning
confidence: 99%
“…Mean difference of eGFRCKD-EPI compared to mGFR was 5±10 mL/min/1.73m 2 for Roche Gen2 and 5±11 mL/min/1.73m 2 for Siemens. There was no significant difference in bias, precision or accuracy between Siemens and Gen 2 for eGFRCKD-EPI as compared to mGFR (Table 3) For the CAPA equation, mean eGFR values were 67±19 mL/min/1.73m 2 with Roche Gen2 and 66±19 mL/min/1.73m 2 with Siemens.…”
Section: Performance Of Estimated Gfr Versus Measured Gfr (N=567)mentioning
confidence: 87%
See 3 more Smart Citations