The purpose of this review is to summarize the research on the accuracy of oxygen saturation (spO 2 ) measurements using the Apple Watch (Apple Inc., Cupertino, California). The Medline and Google Scholar databases were searched for papers evaluating the spO 2 measurements of the Apple Watch vs. any kind of ground truth and records were analyzed according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines.The five publications with 973 total patients that met the inclusion criteria all used the Apple Watch Series 6 and described 95% limits of agreement of +/-2.7 to 5.9% spO 2 . However, outliers of up to 15% spO 2 were reported. Only one study had patient-level data uploaded to a public repository. The Apple Watch Series 6 does not show a strong systematic bias compared to conventional, medical-grade pulse oximeters. However, outliers do occur and should not cause concern in otherwise healthy individuals. The impact of race on measurement accuracy should be investigated.