Accuracy of the CoaguChek XS for point-of-care international normalized ratio (INR) measurement in children requiring warfarin

Abstract: Point-of-care INR (POC INR) meters can provide a safe and effective method for monitoring oral vitamin K antagonists (VKAs) in children. Stollery Children's Hospital has a large POC INR meter loan program for children requiring oral VKAs. Our protocol requires that POC INR results be compared to the standard laboratory INR for each child on several consecutive tests to ensure accuracy of CoaguChek XS (Roche Diagnostics, Basel Switzerland) meter. It was the objective of the study to determine the accuracy of th… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
1
1

Citation Types

0
10
0

Year Published

2009
2009
2023
2023

Publication Types

Select...
7
1

Relationship

0
8

Authors

Journals

citations
Cited by 61 publications
(10 citation statements)
references
References 33 publications
0
10
0
Order By: Relevance
“…Thus, high correlations for 2 given methods do not mean that the 2 methods agree [18]. Besides very few other small reports, only our study and the one by Baumann et al [10] have assessed agreement between venous and capillary INR measurements in a substantial number of children using the Bland-Altman analysis. While both studies indicate that capillary INR measurements using the portable PT monitor are very accurate, the negative values in our study show that the capillary INR was on average lower than the venous INR.…”
Section: Discussionmentioning
confidence: 81%
See 1 more Smart Citation
“…Thus, high correlations for 2 given methods do not mean that the 2 methods agree [18]. Besides very few other small reports, only our study and the one by Baumann et al [10] have assessed agreement between venous and capillary INR measurements in a substantial number of children using the Bland-Altman analysis. While both studies indicate that capillary INR measurements using the portable PT monitor are very accurate, the negative values in our study show that the capillary INR was on average lower than the venous INR.…”
Section: Discussionmentioning
confidence: 81%
“…Unfortunately, only few studies evaluating capillary INR monitoring in children have been conducted so far [10,11,12,13,14,15,16]. In particular, very few pediatric data are available on the newer PT monitor, CoaguChek XS.…”
Section: Introductionmentioning
confidence: 99%
“…The safety and reliability of CoaguChek S and CoaguChek XS have been demonstrated in several studies in both adults and children. [43][44][45][46][47][48][49][50] A number of studies have also compared the performance of CoaguChek S with that of CoaguChek XS in relation to conventional INR measurement. Even though a good agreement between the two CoaguChek models and conventional laboratory-based results has been demonstrated, CoaguChek XS has shown more accurate and precise results than its precursor in both adults and children, especially for higher INR values (> 3.5).…”
Section: Coaguchek Xs Versus Coaguchek Smentioning
confidence: 99%
“…INR results that are below the lower limit of the therapeutic range confer an increased risk of thrombosis in patients, whereas INR results above the upper therapeutic limit confer an increased risk of bleeding events. Many families of children on warfarin have been supported to perform INR self-testing on a point-of care (POC) device, following the completion of a comprehensive training program [ [4] , [5] , [6] ]. Home patient self-testing (PST) programs have been evaluated and demonstrated to provide safe, efficacious warfarin management and improve the quality of life for both children and families [ [4] , [5] , [6] , [7] , [8] ].…”
Section: Introductionmentioning
confidence: 99%
“…Many families of children on warfarin have been supported to perform INR self-testing on a point-of care (POC) device, following the completion of a comprehensive training program [ [4] , [5] , [6] ]. Home patient self-testing (PST) programs have been evaluated and demonstrated to provide safe, efficacious warfarin management and improve the quality of life for both children and families [ [4] , [5] , [6] , [7] , [8] ].…”
Section: Introductionmentioning