2024
DOI: 10.1002/rcs.70029
|View full text |Cite
|
Sign up to set email alerts
|

Accuracy of the Imageless Mode of the ROSA Robotic System for Targeted Resection Thickness in Total Knee Arthroplasty: A Prospective, Single Surgeon Case‐Series Study

Zakareya Gamie,
Eustathios Kenanidis,
Georgios Douvlis
et al.

Abstract: BackgroundWe investigated the accuracy of targeted resection thickness in patients undergoing primary Total Knee Arthroplasty (TKA) using the ROSA robotic system.MethodsCalliper measurements of the distal femur (DF), proximal tibia (PT), and posterior condyles (PC) were taken in 44 patients from June 2023 to January 2024.ResultsPlanned and actual resection depth difference was 0.67 mm ± 0.6 mm (mean ± SD) (p = 0.217) and 0.94 mm ± 1.15 mm (p = 0.4) for medial and lateral DF, 0.93 mm ± 0.81 mm (p = 0.001) and 0… Show more