Background
Microsurgical clipping is a delicate neurosurgical procedure used to treat complex Unruptured Intracranial Aneurysms (UIAs) whose outcome is dependent on surgeon’s experience. Simulations are emerging as excellent complements to standard training, but their adoption is limited by the realism they provide. The aim of this study was to develop and validate a microsurgical clipping simulator platform.
Methods
Physical and holographic simulators of UIA clipping have been developed. The physical phantom consisted of a 3D printed hard skull and five (
n
= 5) rapidly interchangeable, perfused and fluorescence compatible 3D printed aneurysm silicone phantoms. The holographic clipping simulation included a real-time finite-element-model of the aneurysm sac, allowing interaction with a virtual clip and its occlusion. Validity, usability, usefulness and applications of the simulators have been assessed through clinical scores for aneurysm occlusion and a questionnaire study involving 14 neurosurgical residents (R) and specialists (S) for both the physical (
p
) and holographic (
h
) simulators by scores going from 1 (very poor) to 5 (excellent).
Results
The physical simulator allowed to replicate successfully and accurately the patient-specific anatomy. UIA phantoms were manufactured with an average dimensional deviation from design of 0.096 mm and a dome thickness of 0.41 ± 0.11 mm. The holographic simulation executed at 25–50 fps allowing to gain unique insights on the anatomy and testing of the application of several clips without manufacturing costs. Aneurysm closure in the physical model evaluated by fluorescence simulation and post-operative CT revealed Raymond 1 (full) occlusion respectively in 68.89% and 73.33% of the cases. For both the simulators content validity, construct validity, usability and usefulness have been observed, with the highest scores observed in clip selection usefulness R
p
=4.78, S
p
=5.00 and R
h
=4.00, S
h
=5.00 for the printed and holographic simulators.
Conclusions
Both the physical and the holographic simulators were validated and resulted usable and useful in selecting valid clips and discarding unsuitable ones. Thus, they represent ideal platforms for realistic patient-specific simulation-based training of neurosurgical residents and hold the potential for further applications in preoperative planning.
Supplementary Information
The online version contains supplementary material available at 10.1186/s41205-024-00235-w.