1999
DOI: 10.1007/s005860050171
|View full text |Cite
|
Sign up to set email alerts
|

Adolescent idiopathic thoracic scoliosis: apical correction with specialized pedicle hooks

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

0
4
1
1

Year Published

2000
2000
2021
2021

Publication Types

Select...
8

Relationship

1
7

Authors

Journals

citations
Cited by 17 publications
(6 citation statements)
references
References 22 publications
0
4
1
1
Order By: Relevance
“…The average load to failure of the pedicle screws was 1,046.8 ± 173.1 N (n = 10) and 236.2 ± 39.8 N for the SPHD (P < 0.001). The average BMC of the vertebrae was 0.735 ± 0.060 g/cm 2 . The pull-out strength of the pedicle screws showed greater correlation with the BMC (R 2 = 0.4811) compared to the pull-out strength of the SPHD and the BMC (R 2 = 0.1343).…”
Section: Resultsmentioning
confidence: 99%
See 1 more Smart Citation
“…The average load to failure of the pedicle screws was 1,046.8 ± 173.1 N (n = 10) and 236.2 ± 39.8 N for the SPHD (P < 0.001). The average BMC of the vertebrae was 0.735 ± 0.060 g/cm 2 . The pull-out strength of the pedicle screws showed greater correlation with the BMC (R 2 = 0.4811) compared to the pull-out strength of the SPHD and the BMC (R 2 = 0.1343).…”
Section: Resultsmentioning
confidence: 99%
“…However, these instruments disturb the surgical field and tend to dislodge the hook as a result of the long lever arm. Augmentation of the pedicle hook to the lamina and pedicle by a small screw has been proposed for pedicle hook fixation [1,2,[5][6][7]. These special pedicle hooks solve these disadvantages of regular hooks partially, however, fixation of a small screw in the same pedicle that is used for the pedicle hook fixation weakens the pedicle and increases the risk of pedicle fracture [6,10].…”
Section: Fixation Methods Fixation Strength (N)mentioning
confidence: 99%
“…It permits the use of rod translation instead of rod rotation; it also allows for secure pedicle hooks with fixation screws [17]. The use of pedicle screw instrumentation in the lumbar and low thoracic spine as a foundation was a natural evolution in segmental instrumentation.…”
Section: Hybrid Constructs With Segmental Translation (Pedicle Screwsmentioning
confidence: 99%
“…Ariet et al [24] berichteten jedoch über Korrekturen der Scheitelwirbeltranslation von 70 % unter Verwendung von modifizierten Pedikelhaken. Das Ausmaû der Korrektur der Scheitelwirbeltranslation ist zum einen abhängig von der Anzahl der implantierten Anker auf der Konkavität [11], zum anderen von der Ausrissfestigkeit des Implantates.…”
Section: Ergebnisseunclassified